Fatigue-Induced Cole Electrical Impedance Model Changes of Biceps Tissue Bioimpedance

Loading...
Thumbnail Image
Date
2018
Journal Title
Journal ISSN
Volume Title
Publisher
MDPI
Abstract

Bioimpedance, or the electrical impedance of biological tissues, describes the passive electrical properties of these materials. To simplify bioimpedance datasets, fractional-order equivalent circuit presentations are often used, with the Cole-impedance model being one of the most widely used fractional-order circuits for this purpose. In this work, bioimpedance measurements from 10 kHz to 100 kHz were collected from participants biceps tissues immediately prior and immediately post completion of a fatiguing exercise protocol. The Cole-impedance parameters that best fit these datasets were determined using numerical optimization procedures, with relative errors of within approximately +/- 0.5% and +/- 2% for the simulated resistance and reactance compared to the experimental data. Comparison between the pre and post fatigue Cole-impedance parameters shows that the R-infinity, R-1, and f(p) components exhibited statistically significant mean differences as a result of the fatigue induced changes in the study participants.

Description
Keywords
bioimpedance, Cole-impedance model, fractional-order model, tissue, fatigue, FRACTIONAL CALCULUS, PARAMETERS, ALGORITHMS, Mathematics, Interdisciplinary Applications, Mathematics
Citation
Freeborn, T.; Fu, B. Fatigue-Induced Cole Electrical Impedance Model Changes of Biceps Tissue Bioimpedance. Fractal Fract. 2018, 2(4), 27; https://doi.org/10.3390/fractalfract2040027.